PATIENT WARMER SYSTEM

Category: >Plastic Surgery Equipment

Description

        SPECIFICATION

  •  TEMPERATURE SETTING DESIGNED TO OFFSET PATIENTHYPOTHERMIA
  •  HIGH- 43°
  •  MED-38°
  •  LOW- 32°
  •  AMBIENT MODE-HEATER WILL REMAIN OFF, ONLY BLOWER WILL WORK MPROVE RECOVERY TIME
  •  MINIMIZES POST SURGICAL COMPLICATIONS
  •  FULL BODY FLEXIBLE BLANKET MADE BY SOFT BUT STRONG QUILTED MATERIAL TO
  •  PROMOTE UNIFORM AIR FLOW AND COMFORT TO THE PATIENT BODY
  •  AUDIBLE ALARM WITH LED INDICATIONFOR THE FOLLOWING
  • HIGH TEMP- DO NOT USE PATIENT WARMER
  • LOW TEMP- HEATER TO BE CHECKED
  • SENSOR OPEN- TEMP SENSOR TO BE CHECKED
  • FILTER CHANGE-FILTER LIFE EXPECTANCY IS OVER, REPLACE WITH NEW FILTER HEPA FILTER
  • HIGH EFFICIENCY PARTICULATE AIR FILTER 99.97% EFFICIENT AT 0.3-MICRON PARTICLE
  • SIZE H-16” x W-10” x D-8.5” WIEGHT -8.60 KG
  •  HEATER SIZE 1000 WATTS
  •  WARM UP TIME – AIR TEMPERATURE AT NOZZLE EXCEEDS 30° C IN APPROXIMATELY 30
  • SECONDS
  •  NOISE LEVEL- LESS THAN 55db

      ACCESSORIES

  • WARMER MACHINE
  • BLANKET
  • POWER CORD